Subject(s) - chromatography , chemistry , immunoassay , urine , high performance liquid chromatography , buprenorphine , codeine , dextrorphan , dextromethorphan , morphine , opioid , pharmacology , medicine , biochemistry , receptor , antibody , immunology
A solid-phase enzyme immunoassay involving microtiter plates was proposed by Microgenics to screen buprenorphine in urine. The intra-assay precision at 10 ng/mL was 7.7% (coefficient of variation). The immunoassay was determined to have no cross-reactivity with codeine, dihydrocodeine, morphine, ethylmorphine, 6-monoacetylmorphine, methadone, pholcodine, propoxyphene, dextromoramide, and dextromethorphan at 1 and 10 mg/L. A low cross-reactivity (3% at 1 ng/mL) was observed at low concentrations of norbuprenorphine. After comparing this new immunological test (Singlestep ELISA) for 76 urine specimens with our validated high-performance liquid chromatography-electrospray mass spectrometry (HPLC-ES-MS) procedure, an optimum cutoff concentration of 2 ng/mL was determined for the kit. At this cutoff, the screening assay was able to determine more than 90% of true results with 43.4% true positives and 48.7% true negatives. Four positive urines (5.3%) were not confirmed by HPLC-ES-MS. In only one case, the negative urine test was confirmed as positive by HPLC-ES-MS (buprenorphine: 62.5 ng/mL). Buprenorphine concentrations determined by HPLC-ES-MS ranged from 1.2 to 1052 ng/mL. Of the four potential adulterants (hypochloride 50 mL/L, sodium nitrite 50 g/L, liquid soap 50 mL/L, and sodium chloride 50 g/L) that might be added to a positive urine specimen, none were able to cause a false-negative response by the immunoassay. The results of this study support the concept that the Singlestep ELISA for buprenorphine determination in urine should be considered as a new, valided screening procedure.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom