Injuries to the superior vena cava (SVC) during transvenous lead extraction (TLE) procedures are a rare but life-threatening complication. The Bridge Occlusion Balloon (BOB) is specifically designed for temporary SVC occlusion in TLE procedures. We report the first case of a 27-year-old man using the BOB as a safety net in a high-risk TLE procedure. This patient, with a congenitally corrected transposition of the great arteries and a third-degree atrioventricular block, presented with 4 dysfunctional pacemaker leads, venous stenosis and the necessity for a new pacemaker system. The leads were implanted for 10 and 19 years. The BOB was placed with a radiopaque marker at the cavoatrial junction and was inflated with 46 ml of an 80/20 saline/contrast agent mixture. An angiography was performed to confirm SVC occlusion. With the deflated balloon in place, the TLE procedure with laser and mechanical sheaths was performed. Successful extraction of 2 dysfunctional leads, as well as venous recanalization, for the new right atrial and right ventricular lead implantation was achieved. We have shown the feasibility of using powered extraction sheaths with a deflated BOB in place. This allows for immediate balloon inflation, in case of an SVC perforation.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom