Twiddling in cardiac resynchronization therapy: 'when length matters'
Author(s) -
Miguel Á. Arias,
Alberto Puchol,
Marta Pachón,
Eduardo Castellanos,
Luis Rodrı́guez Padial
Publication year - 2008
Publication title -
ep europace
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 2.119
H-Index - 102
eISSN - 1532-2092
pISSN - 1099-5129
DOI - 10.1093/europace/eun394
Subject(s) - medicine , cardiac resynchronization therapy , cardiology , radiography , heart failure , surgery , ejection fraction
We report on a 75-year-old woman who successfully underwent cardiac resynchronization therapy. Two weeks after implantation, right continuous pectoral stimulation appeared. Retraction and dislodgement of the right atrial and right ventricular leads as well as coiling of the three leads around the device generator were observed on chest radiography, but, surprisingly, the left ventricular lead remained unaffected due to its major length to facilitate the manoeuvers of implantation.