Technical performance evaluation of BM/Hitachi 747-200 serum fructosamine assay
Author(s) -
MeiChun Lin,
Carolyn Hoke,
Bruce Ettinger,
Rosie Coyne
Publication year - 1996
Publication title -
clinical chemistry
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.705
H-Index - 218
eISSN - 1530-8561
pISSN - 0009-9147
DOI - 10.1093/clinchem/42.2.244
Subject(s) - fructosamine , hemoglobin , endocrinology , medicine , chemistry , diabetes mellitus , chromatography
The technical performance of serum fructosamine assay as a short-term index of diabetic control was evaluated by using the BM/Hitachi 747-200 Automatic Analyzer (Boehringer Mannheim Corp). Intra- and interassay precisions (CV) were <2%. Linearity was confirmed up to 1000 micromol/L. Lipemia, produced artificially or from hypertriglyceridemia in vivo, did not interfere with measurement of serum fructosamine concentration. However, hemoglobin (Hb) interfered with measurement substantially at concentrations >15.5 micromol/L, and bilirubin interfered substantially at concentrations >68.4 micromol/L. A normal reference range of up to 282 micromol/L (mean +/- 2SD = 242 +/- 40) was derived from results obtained from 228 nondiabetic subjects. In 222 diabetic subjects, serum fructosamine concentration correlated well with fasting glucose concentration (r = 0.74) and with Hb A1c concentration (r = 0.80). This automated fructosamine assay has the advantages of technical simplicity, low cost, and reduced analytical time compared with the Hb A1c method.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom