Validation of an Ultra-High Performance Liquid Chromatography Tandem Mass Spectrometric Method for Quantifying Uracil and 5,6-Dihydrouracil in Human Plasma
Author(s) -
François Coudoré,
Denis Roche,
Sandrine Lefeuvre,
Delphine Faussot,
Eliane M. Billaud,
MarieAnne Loriot,
Philippe Beaune
Publication year - 2012
Publication title -
journal of chromatographic science
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.362
H-Index - 56
eISSN - 1945-239X
pISSN - 0021-9665
DOI - 10.1093/chromsci/bms085
Subject(s) - chemistry , chromatography , high performance liquid chromatography , selected reaction monitoring , calibration curve , mass spectrometry , triple quadrupole mass spectrometer , electrospray ionization , uracil , analyte , analytical chemistry (journal) , detection limit , tandem mass spectrometry , dna , biochemistry
A simple, rapid, sensitive and specific ultra-high-performance liquid chromatography-tandem mass spectrometry method (Waters UPLC-MS-MS) is developed and validated for the quantification of uracil (U) and 5,6-dihydrouracil (UH2) levels in human plasma. Analytes are extracted using ethyl acetate and isopropanol after deproteination, and separated by high-performance liquid chromatography (HPLC) (Acquity UPLC BEH C18 column) in a binary mobile phase system under gradient elution conditions at a flow rate of 0.4 mL/min. 5-Bromo-uracil (UBr) is used as the internal standard. The detection is performed on a triple-quadrupole mass spectrometer via electrospray positive ionization. Multiple reaction monitoring mode using the transitions m/z 112.82 → 70.05, m/z 114.88 → 55.04 and m/z 190.83 → 117.86 is used to quantify U, UH2 and UBr, respectively. The method is linear in the concentration range of 0.625-160.0 ng/mL. The total run time is 4.5 min per injection. Nine-point calibration curve and four-points quality controls are used. Excellent linearity and precision are observed with correlation coefficient (r(2)) > 0.9999. The intra-batch and inter-batch precisions are ≤ 7.3% and ≤ 8.6%, and accuracy is ≤ 17%. The developed method is shown to be suitable for routine quantitative determination of U, UH2 and 5,6-dihydrouracil-to-uracil ratio in clinical practice.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom