Evaluation of a Capillary Gas Chromatographic Impurity Test Procedure for 4-Hexylaniline
Author(s) -
Clayton B’Hymer
Publication year - 2005
Publication title -
journal of chromatographic science
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.362
H-Index - 56
eISSN - 1945-239X
pISSN - 0021-9665
DOI - 10.1093/chromsci/43.5.267
Subject(s) - chromatography , chemistry , impurity , flame ionization detector , repeatability , capillary action , gas chromatography , isothermal process , methanol , analytical chemistry (journal) , polyethylene glycol , organic chemistry , materials science , physics , composite material , thermodynamics
A capillary gas chromatographic test procedure for the detection and quantitation of impurities in the bulk intermediate, 4-hexylaniline, is evaluated and found to be accurate and precise. 4-Hexylaniline is dissolved in methanol and chromatographed isothermally at a temperature of 195 degrees C on a 60-m x 0.32-mm 85% polyethylene glycol-15% dimethylsilicone blend (DX-4) film column. A flame ionization detector is used, and the impurities in the parent compound are estimated from peak areas on a percent basis compared with the area of the parent peak in the chromatogram. Response factors are determined for the known impurities. Validation of this test method includes a recovery study of known impurity spiked samples fortified in the range of 0.1-1% (w/w). A repeatability study is performed, consisting of the analysis of two different synthetic batch lots of 4-hexylaniline analyzed over three experimental run days using two chromatographic columns of different manufacturing lots. These data and other aspects of this test procedure are discussed.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom