Gas Chromatography Problem Solving and Troubleshooting
Author(s) -
D. Rood
Publication year - 1998
Publication title -
journal of chromatographic science
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.362
H-Index - 56
eISSN - 1945-239X
pISSN - 0021-9665
DOI - 10.1093/chromsci/36.2.109
Subject(s) - chemistry , troubleshooting , chromatography , gas chromatography , reliability engineering , engineering
With only a few exceptions, capillary columns are not directional, and their performance is the same regardless of the installation direction. Some types of columns are more prone to directional differences. Long-length, thick-film, or small-diameter columns are more likely to suffer from a directional difference. A directional difference is even more likely if a column has more than one of these characteristics. If a directional difference occurs, it indicates that the front portion of the column is different than the back portion. This difference is in the stationary phase film thickness; in other words, the film thickness is different at the two column ends. Usually, the difference in film thickness occurs as a gradient along the column length. In other words, the film gradually becomes thicker or thinner along the length of the column. One way to illustrate this occurrence is to cut the column in half and compare the retention of the two halves. Figure 1 shows a test mixture chromatogram for a 60-m column exhibiting a directional difference. Figure 2 shows the same test mixture for each of the 30-m halves. Pentadecane (peak 8) retention times of 15.99 and 11.40 min are obtained for the two halves. A retention time difference of less than 0.5 min would normally be expected for this type of column. Comparing pentadecane retention factors (k) for all three columns further illustrates the film thickness differences. If the film thickness was uniform throughout the column, the same retention factor would be obtained for the original column and the two halves. The retention factor for the original 60-m column is 8.06 (Figure 1). Retention factors of 9.12 and 6.50 were obtained for the two halves (Figures 2A and 2B). These retention factors indicate that the average film thickness of the two columns are 0.8 and 0.55 μm, respectively. Sometimes, a longer column is cut into shorter lengths. This is done to save money, because longer columns cost less per meter than shorter columns. Although this practice usually works, the occasional directional difference in columns may result in the shorter columns having significantly different retention characteristics. Gas Chromatography Problem Solving and Troubleshooting
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom