Impact of hemodialysis on dual X-ray absorptiometry, bioelectrical impedance measurements, and anthropometry
Author(s) -
Bo Abrahamsen,
TB Hansen,
I M Høgsberg,
F B Pedersen,
Henning BeckNielsen
Publication year - 1996
Publication title -
american journal of clinical nutrition
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 2.608
H-Index - 336
eISSN - 1938-3207
pISSN - 0002-9165
DOI - 10.1093/ajcn/63.1.80
Subject(s) - bioelectrical impedance analysis , anthropometry , hemodialysis , medicine , dual energy x ray absorptiometry , bone mineral , nuclear medicine , lean body mass , body water , dialysis , urology , surgery , body weight , body mass index , osteoporosis
Dual X-ray absorptiometry (DXA) performs noninvasive assessment of bone and soft tissue with high precision. However, soft tissue algorithms assume that 73.2% of the lean body mass is water, a potential source of error in fluid retention. We evaluated DXA (model QDR-2000; Hologic Inc, Waltham, MA), bioelectrical impedance analysis (BIA), and simple anthropometry in 19 patients (9 women and 10 men, mean age 46 y) before and after hemodialysis, removing 0.9-4.3 L (x: 2.8L) of ultrafiltrate. The reduction in fat-free mass (FFM) measured by DXA was highly correlated with the ultrafiltrate, as determined by the reduction in gravimetric weight (r = 0.975, P < 0.0001; SEE: 233 g), whereas BIA was considerably less accurate in assessing FFM reductions (r = 0.66, P < 0.01; SEE: 757 g). Lumbar bone mineral density (BMD) was unaffected by dialysis, as were whole-body fat and BMD. Whole-body bone mineral content, however, was estimated to be 0.6% lower after dialysis. None of the simple anthropometric measurements correlated significantly with the reduction in FFM. In an unmodified clinical setting, DXA appears to be superior to other simple noninvasive methods for determining body composition, particularly when the emphasis is on repeated measurements.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom