An enhanced rheometer inertia correction procedure (ERIC) for the study of gelling systems using combined motor-transducer rheometers
Author(s) -
Ralph E. Hudson,
A. Holder,
Karl Hawkins,
P. R. Williams,
Daniel J. Curtis
Publication year - 2017
Publication title -
physics of fluids
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.188
H-Index - 180
eISSN - 1089-7666
pISSN - 1070-6631
DOI - 10.1063/1.4993308
Subject(s) - rheometer , viscoelasticity , rheology , transducer , shear modulus , gel point , physics , relaxation (psychology) , inertia , biological system , mechanics , materials science , thermodynamics , acoustics , classical mechanics , psychology , social psychology , biology
The rheological characterisation of viscoelastic materials undergoing a sol-gel transition at the Gel Point (GP) has important applications in a wide range of industrial, biological, and clinical environments and can provide information regarding both kinetic and microstructural aspects of gelation. The most rigorous basis for identifying the GP involves exploiting the frequency dependence of the real and imaginary parts of the complex shear modulus of the critical gel (the system at the GP) measured under small amplitude oscillatory shear conditions. This approach to GP identification requires that rheological data be obtained over a range of oscillatory shear frequencies. Such measurements are limited by sample mutation considerations (at low frequencies) and, when experiments are conducted using combined motor-transducer (CMT) rheometers, by instrument inertia considerations (at high frequencies). Together, sample mutation and inertia induced artefacts can lead to significant errors in the determinatio...
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom