Providing radiometric traceability for the calibration home base of DLR by PTB
Author(s) -
D. R. Taubert,
J. Hollandt,
Peter Sperfeld,
S. Pape,
Andreas Höpe,
K.-O. Hauer,
Peter Gege,
Thomas Schwarzmaier,
Karim Lenhard,
Andreas Baumgartner
Publication year - 2013
Publication title -
aip conference proceedings
Language(s) - English
Resource type - Conference proceedings
SCImago Journal Rank - 0.177
H-Index - 75
eISSN - 1551-7616
pISSN - 0094-243X
DOI - 10.1063/1.4804785
Subject(s) - radiance , calibration , irradiance , radiometric calibration , remote sensing , optics , range (aeronautics) , physics , materials science , geology , quantum mechanics , composite material
A dedicated calibration technique was applied for the calibration of the spectral radiance transfer standard (RASTA) of the German Aerospace Center (DLR) at the Physikalisch-Technische Bundesanstalt (PTB), consisting of two independent but complementing calibration procedures to provide redundancy and smallest possible calibration uncertainties. Procedure I included two calibration steps: In a first step the optical radiation source of RASTA, an FEL lamp, was calibrated in terms of its spectral irradiance Eλ(λ) in the wavelength range from 350 nm to 2400 nm using the PTB Spectral Irradiance Calibration Equipment (SPICE), while in a second step the spectral radiance factor β0°:45°(λ) of the RASTA reflection standard was calibrated in a 0°:45°-viewing geometry in the wavelength range from 350 nm to 1700 nm at the robot-based gonioreflectometer facility of PTB. The achieved relative standard uncertainties (k = 1) range from 0.6 % to 3.2 % and 0.1 % to 0.6 % respectively. Procedure II was completely independe...
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom