Analyzing refractive index changes and differential bending in microcantilever arrays
Author(s) -
François Huber,
H.P. Lang,
Martin Hegner,
M. Despont,
Ute Drechsler,
Christoph Gerber
Publication year - 2008
Publication title -
review of scientific instruments
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.605
H-Index - 165
eISSN - 1089-7623
pISSN - 0034-6748
DOI - 10.1063/1.2969912
Subject(s) - refractive index , materials science , curvature , bending , cantilever , optics , laser , refractive index profile , optoelectronics , composite material , physics , geometry , mathematics
A new microcantilever array design is investigated comprising eight flexible microcantilevers introducing two solid bars, enabling to subtract contributions from differences in refractive index in an optical laser read out system. Changes in the refractive index do not contribute undesirably to bending signals at picomolar to micromolar DNA or protein concentrations. However, measurements of samples with high salt concentrations or serum are affected, requiring corrections for refractive index artifacts. Moreover, to obtain a deeper understanding Of molecular stress formation, the differential curvature of cantilevers is analyzed by positioning the laser spots along the surface of the levers during pH experiments