A grounded-load charge amplifier for reducing hysteresis in piezoelectric tube scanners
Author(s) -
Andrew J. Fleming,
S. O. Reza Moheimani
Publication year - 2005
Publication title -
review of scientific instruments
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.605
H-Index - 165
eISSN - 1089-7623
pISSN - 0034-6748
DOI - 10.1063/1.1938952
Subject(s) - amplifier , hysteresis , voltage , charge amplifier , tube (container) , piezoelectricity , materials science , electrical engineering , operational amplifier , charge (physics) , optoelectronics , physics , engineering , condensed matter physics , composite material , cmos , quantum mechanics
In this paper, a charge amplifier adapted for piezoelectric tube scanners is presented. Previous problems involved with the implementation of such amplifiers are resolved to provide dc accurate performance with zero voltage drift. In our experiment, hysteresis was reduced by 89% when compared to a voltage amplifier.