Simple pressure gauge for uranium hexafluoride
Author(s) -
Patricia M. George,
Mark N. Swanson,
J. L. Beauchamp
Publication year - 1979
Publication title -
review of scientific instruments
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.605
H-Index - 165
eISSN - 1089-7623
pISSN - 0034-6748
DOI - 10.1063/1.1135745
Subject(s) - uranium hexafluoride , torr , rhenium , pressure measurement , materials science , ionization , atomic physics , analytical chemistry (journal) , uranium , ion , physics , chemistry , thermodynamics , metallurgy , chromatography , quantum mechanics
A sensitive detector and pressure gauge for uranium hexafluoride in high-vacuum systems is described. Negative surface ionization of UF(6) occurs on ribbon filaments operated at temperatures too low for electron emission to be significant. The ion current measured on a cylindrical collector surrounding the filament assembly varies regularly with UF(6) pressure below 10(-3) Torr. Different filament materials are considered, including rhenium, thoriated tungsten, and platinum. Rhenium is found to be the most satisfactory material for operation of diode emitters as a pressure gauge. Gauge constants (in A Torr(-1)) are derived from comparing negative surface ionization currents with the response of a capacitance manometer and are shown to be independent of temperature within a reasonable operating range. The effects of exposing the rhenium filament to various gases is considered, and it is shown that brief exposure to acetylene substantially improves the operating characteristics of the gauge.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom