Implantable loop recorder undersensing mimicking complete heart block
Author(s) -
Stavros I. Chrysostomakis,
Εmmanuel Simantirakis,
Maria Marketou,
Panos Vardas
Publication year - 2002
Publication title -
ep europace
Language(s) - Uncategorized
Resource type - Journals
SCImago Journal Rank - 2.119
H-Index - 102
eISSN - 1532-2092
pISSN - 1099-5129
DOI - 10.1053/eupc.2002.0229
Subject(s) - medicine , implantable loop recorder , heart block , loop (graph theory) , cardiology , electrocardiography , atrial fibrillation , mathematics , combinatorics
The newer implantable loop recorders recently introduced (Reveal Plus. Medtronic Inc, Minneapolis, MN, U.S.A.). equipped with auto activation capabilities, are expected to expand our diagnostic abilities in the investigation of syncopal episodes. The case presented here is that of a man with pre-syncopal and syncopal episodes, without organic heart disease, in whom the device was auto-activated due to undersensing and the recorded strip imitated complete heart block. Increased attention should be given by the interpreter of the device's recordings to avoid false positive diagnoses that might lead to serious therapeutic actions.