Latest generation of unipolar and bipolar steroid eluting leads Long-term comparison of electrical performance in atrium and ventricles
Author(s) -
D Cornacchia,
M. Fabbri,
Andrea Puglisi,
P V Moracchini,
M. Bernasconi,
M Nastasi,
C. Menozzi,
Giosué Mascioli,
T. Marotta,
Francesco De Seta
Publication year - 2000
Publication title -
ep europace
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 2.119
H-Index - 102
eISSN - 1532-2092
pISSN - 1099-5129
DOI - 10.1053/eupc.2000.0103
Subject(s) - medicine , ventricle , atrium (architecture) , cardiology , ventricular pacing , heart failure , atrial fibrillation
One hundred and seventy-three patients, mean age 74 years permanently paced with 123 atrial (53 unipolar, 70 bipolar) and 143 ventricular (73 unipolar, 70 bipolar) pacing leads were included in this study. The pacing leads were recent generation low surface area steroid eluting leads from one manufacturer: leads with silicone and polyurethane insulation were studied, and they were combined with generations of one pacemaker family from the same manufacturer permitting identical measurements to be made over a follow-up of 2 years. Pacing threshold was measured using pulse duration at a fixed voltage of 1.5 V: peak to peak P and R wave amplitude and pacing impedance at 2.5 V and 0.5 ms were all measured using the manufacturer's standard programmer. Although many significant differences, in the parameters measures, existed between atrium and ventricle and unipolar and bipolar configurations, none was felt to be of clinical significance. These data permit the physician to choose the lead type with regard to sensing performance and long-term lead integrity.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom