Comparative analysis of the wear of titanium/titanium and titanium/zirconia interfaces in implant/abutment assemblies after thermocycling and mechanical loading
Author(s) -
Paulo Almeida,
César L. Silva,
Jorge Lino Alves,
F.S. Silva,
Ramiro C. Martins,
João Sampaio Fernandes
Publication year - 2016
Publication title -
revista portuguesa de estomatologia medicina dentária e cirurgia maxilofacial
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.144
H-Index - 8
eISSN - 1647-6700
pISSN - 1646-2890
DOI - 10.1016/j.rpemd.2016.07.002
Subject(s) - titanium , cubic zirconia , abutment , materials science , profilometer , implant , scanning electron microscope , dentistry , dental abutments , composite material , metallurgy , structural engineering , medicine , ceramic , surface finish , engineering , surgery
ObjectivesDue to its hardness, zirconia abutments may damage the titanium of the implant's connection during its clinical use. This study aimed to assess the wear of the seating platforms of externally hexed titanium implants when connected to zirconia abutments comparing to titanium abutments, after thermocycling and mechanical loading (TCML).MethodsSix BNT® S4 external connection implants (Phibo® Dental Solutions, Barcelona, Spain) were selected and divided into two groups (n=3): TiCE, screwed to titanium abutments and ZrCE, screwed to zirconia abutments. The samples underwent thermocycling (5000 cycles; 5–55°C) and mechanical loading (1.2×106 cycles; 88.8N; 4Hz). Before and after TCML, the seating platforms of the implants were analyzed by 3D profilometry in two areas to measure their topography based on the superficial analysis parameters (Sa and Sz) parameters, and were studied by scanning electron microscopy (SEM). Data were statistically analyzed by Mann–Whitney test (p<0.05).ResultsNo statistically significant differences were observed in the Sa and Sz values between the implants at initial state and after TCML (p=0.573>0.05 and p=0.059). The abutment's material (titanium/zirconia) did not statistically significantly influence the Sa and Sz values after TCML (p=0.886>0.05 and p=0.200, respectively). However, the SEM analysis reveals a mild wear in some vertices of the hex when connected to zirconia abutments.ConclusionsAfter simulating five years of clinical use, the externally hexed implants, analyzed by 3D profilometry, showed similar wear patterns in the corresponding seating platforms when connected to zirconia or titanium abutments. The SEM images showed zirconia particles being transferred to the implant, which requires further study
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom