Failure to deliver a shock in a dual-chamber implantable cardioverter-defibrillator: A case report
Author(s) -
Pedro A. Sousa,
Rui Candeias,
Ilídio de Jesus
Publication year - 2015
Publication title -
revista portuguesa de cardiologia
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.266
H-Index - 26
eISSN - 2174-2030
pISSN - 0870-2551
DOI - 10.1016/j.repc.2014.12.008
Subject(s) - medicine , cardiology , implantable cardioverter defibrillator , ventricular tachycardia , atrial fibrillation , supraventricular tachycardia , shock (circulatory) , implant , heart failure , tachycardia , surgery
Inappropriate implantable cardioverter-defibrillator (ICD) therapies due to supraventricular tachyarrhythmia (SVT) are a common problem. The authors report this case to warn of a possible detection problem and subsequent failure of deliver appropriate therapy in patients with atrial fibrillation (AF) and a dual-chamber ICD using the PARAD+ algorithm. To our knowledge this is the first reported case of failure to deliver a shock in a dual-chamber ICD due to the PARAD+ algorithm.