Detection of Natural Cr(VI) with Computer Screen Photo-assisted Technology
Author(s) -
Marco Santonico,
Andrea Scozzari,
G. Brozzo,
Luigi Marini,
A. D’Amico,
Daniel Filippini,
Ingemar Lundström,
Corrado Di Natale
Publication year - 2009
Publication title -
procedia chemistry
Language(s) - English
Resource type - Journals
ISSN - 1876-6196
DOI - 10.1016/j.proche.2009.07.079
Subject(s) - natural (archaeology) , materials science , computer science , geology , paleontology
The combined use of a computer screen and a webcam has been demonstrated in several cases to provide a platform for measuring optical properties of liquid and solid samples. Here the method is utilized to measure the absorbance change in an analytical procedure aimed at determining the amount of hexavalent chromium (Cr(VI)) in water. The method calibrated on standard solutions was tested on-field on naturally contaminated spring waters. Results show a sufficient resolution to measure Cr(VI) concentration down to 5 ppb. World Health Organization recommends a maximum concentration of 0.05 mg/l (50 ppb) for chromium (as hexavalent) in drinking waters