z-logo
open-access-imgOpen Access
Laceration of atrial septum during balloon sizing of atrial septal defect
Author(s) -
S Sivasankaran,
S Harikrishnan,
Njanoor Narayanan,
Tharakan Jaganmohan
Publication year - 2006
Publication title -
european journal of echocardiography
Language(s) - English
Resource type - Journals
eISSN - 1525-2167
pISSN - 1532-2114
DOI - 10.1016/j.euje.2006.10.007
Subject(s) - medicine , atrial septum , balloon , cardiology
We read with great interest the recently published case report by Alsaileek et al.1 in the European Journal of Echocardiography on laceration of atrial septum during balloon sizing. We had already reported about tear of atrial septum during balloon sizing of atrial septal defect (ASD) in a 20-year-old female.2 Balloon sizing was attempted with Amplatzer sizing balloon (Amplatzer sizing Balloon, AGA medical corporation, 682 Mendelssohn Avenue, Golden Valley, MN-55427) (length, 24mm; maximum volume, 30ml). At partial inflation a good indentation was obtained which slowly disappeared without further increase in the balloon volume. Trans-thoracic echocardiography (TTE) followed by trans-esophageal echocardiography (TEE) revealed a tear in the inferior margin …

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom