Author(s) -
Mathieu Vergnat,
Roland Hénaine,
Mārtiņš Kalējs,
Sandra Bommeli,
Enrico Ferrari,
JeanFrançois Obadia,
Ludwig Karl von Segesser
Publication year - 2009
Publication title -
european journal of cardio-thoracic surgery
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.303
H-Index - 133
eISSN - 1873-734X
pISSN - 1010-7940
DOI - 10.1016/j.ejcts.2009.01.052
Subject(s) - stent , medicine , aortic valve , cardiology , ventricle , balloon , hemodynamics , balloon dilation , fixation (population genetics) , population , environmental health
Balloon-expandable stent valves require flow reduction during implantation (rapid pacing). The present study was designed to compare a self-expanding stent valve with annular fixation versus a balloon-expandable stent valve.Implantation of a new self-expanding stent valve with annular fixation (Symetis, Lausanne, Switzerland) was assessed versus balloon-expandable stent valve, in a modified Dynatek Dalta pulse duplicator (sealed port access to the ventricle for transapical route simulation), interfaced with a computer for digital readout, carrying a 25 mm porcine aortic valve. The cardiovascular simulator was programmed to mimic an elderly woman with aortic stenosis: 120/85 mmHg aortic pressure, 60 strokes/min (66.5 ml), 35% systole (2.8 l/min).A total of 450 cardiac cycles was analysed. Stepwise expansion of the self-expanding stent valve with annular fixation (balloon-expandable stent valve) resulted in systolic ventricular increase from 120 to 121 mmHg (126 to 830+/-76 mmHg)*, and left ventricular outflow obstruction with mean transvalvular gradient of 11+/-1.5 mmHg (366+/-202 mmHg)*, systolic aortic pressure dropped distal to the valve from 121 to 64.5+/-2 mmHg (123 to 55+/-30 mmHg) N.S., and output collapsed to 1.9+/-0.06 l/min (0.71+/-0.37 l/min* (before complete obstruction)). No valve migration occurred in either group. (*=p<0.05).Implantation of this new self-expanding stent valve with annular fixation has little impact on haemodynamics and has the potential for working heart implantation in vivo. Flow reduction (rapid pacing) is not necessary.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom