Functional biomechanical performance of a novel anatomically shaped polycarbonate urethane total meniscus replacement
Author(s) -
Vrancken A. C. T.,
Eggermont F.,
Tienen T. G.,
Hannink G.,
Buma P.,
Janssen D.,
Verdonschot N.
Publication year - 2016
Publication title -
knee surgery, sports traumatology, arthroscopy
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.806
H-Index - 125
eISSN - 1433-7347
pISSN - 0942-2056
DOI - 10.1007/s00167-015-3632-6
Subject(s) - implant , cadaveric spasm , medial meniscus , meniscus , biomechanics , medicine , transplantation , fibrous joint , femur , fixation (population genetics) , orthodontics , anatomy , surgery , osteoarthritis , incidence (geometry) , population , physics , alternative medicine , environmental health , pathology , optics
Purpose To evaluate the functional biomechanical performance of a novel anatomically shaped, polycarbonate urethane total meniscus implant. Methods Five human cadaveric knees were flexed between 0° and 90° under compressive loads mimicking a squat movement. Anteroposterior (AP) laxity tests were performed in 30° and 90° flexion. Meniscal kinematics and knee laxity were quantified using roentgen stereophotogrammetric analysis. Tibial cartilage contact mechanics were determined in 90° flexion. Measurements were repeated for the native medial meniscus, the implant, after total medial meniscectomy and allograft transplantation. Results The implant and allograft displayed increased posterior and medial displacements compared to the native meniscus, yet no differences were found between the implant and allograft. Meniscal condition did not affect rotational laxity. Compared to the native joint, AP laxity for the implant was increased in 30° flexion, but not in 90°. The implant reduced the mean contact pressure compared to meniscectomy but could not restore contact pressures to native meniscus levels. Compared to the native meniscus, the implant significantly increased the peak pressure, while the contact area was reduced. Contact mechanics of the implant and allograft were never statistically different. Conclusions Biomechanical performance was similar for the implant and allograft. However, both meniscal replacements could not restore outcomes to native meniscus levels or sufficiently improve outcomes after meniscectomy. This was presumably caused by the mobility allowed by the suture‐only horn fixation. The similarity of implant and allograft performance suggests that the novel implant has the biomechanical potential to serve as an alternative to meniscal allograft transplantation.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here