Hammock effect and complete cusp prolapse: Rare mechanisms of Melody valve failure demonstrated by intracardiac echocardiography
Author(s) -
Simone Jhaveri,
Lourdes Prieto,
Patcharapong Suntharos
Publication year - 2019
Publication title -
annals of pediatric cardiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.292
H-Index - 19
eISSN - 0974-2069
pISSN - 0974-5149
DOI - 10.4103/apc.apc_143_18
Subject(s) - medicine , cardiology , intracardiac injection , endocarditis , cusp (singularity) , regurgitation (circulation) , heart failure , etiology , pulmonary valve , geometry , mathematics
Transcatheter pulmonary valve replacement using Melody valve (Medtronic, Minneapolis MN) has significantly increased in the recent decades. Melody valve failures, although rare, can be problematic and require re-intervention. Through intracardiac echocardiography, we present two patients who each had a rare etiology for dysfunction of their Melody valve. Hammock effect, wherein the valve does not oppose the stent and complete cusp failure causing severe regurgitation have not been previously described as causes of Melody valve failure in the absence of endocarditis. Awareness and knowledge of these mechanisms is pivotal in the management of this patient population.