Analysis of small field percent depth dose and profiles: Comparison of measurements with various detectors and effects of detector orientation with different jaw settings
Author(s) -
Henry Finlay Godson,
M Ravikumar,
S Sathiyan,
K M Ganesh,
Y Retna Ponmalar,
Varatharaj Chandraraj
Publication year - 2016
Publication title -
journal of medical physics/journal of medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.292
H-Index - 24
eISSN - 1998-3913
pISSN - 0971-6203
DOI - 10.4103/0971-6203.177284
Subject(s) - multileaf collimator , detector , flatness (cosmology) , orientation (vector space) , collimator , optics , perpendicular , ionization chamber , physics , dosimetry , linear particle accelerator , field (mathematics) , photon , beam (structure) , diode , nuclear medicine , mathematics , geometry , ion , optoelectronics , medicine , cosmology , quantum mechanics , pure mathematics , ionization
The advent of modern technologies in radiotherapy poses an increased challenge in the determination of dosimetric parameters of small fields that exhibit a high degree of uncertainty. Percent depth dose and beam profiles were acquired using different detectors in two different orientations. The parameters such as relative surface dose (D S), depth of dose maximum (D max), percentage dose at 10 cm (D 10), penumbral width, flatness, and symmetry were evaluated with different detectors. The dosimetric data were acquired for fields defined by jaws alone, multileaf collimator (MLC) alone, and by MLC while the jaws were positioned at 0, 0.25, 0.5, and 1.0 cm away from MLC leaf-end using a Varian linear accelerator with 6 MV photon beam. The accuracy in the measurement of dosimetric parameters with various detectors for three different field definitions was evaluated. The relative D S(38.1%) with photon field diode in parallel orientation was higher than electron field diode (EFD) (27.9%) values for 1 cm ×1 cm field. An overestimation of 5.7% and 8.6% in D 10 depth were observed for 1 cm ×1 cm field with RK ion chamber in parallel and perpendicular orientation, respectively, for the fields defined by MLC while jaw positioned at the edge of the field when compared to EFD values in parallel orientation. For this field definition, the in-plane penumbral widths obtained with ion chamber in parallel and perpendicular orientation were 3.9 mm, 5.6 mm for 1 cm ×1 cm field, respectively. Among all detectors used in the study, the unshielded diodes were found to be an appropriate choice of detector for the measurement of beam parameters in small fields.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here