SU‐F‐T‐318: Sensitivity and Stability of OSLDs with Filled Deep Electron/hole Traps Under Pre‐Irradiation and Bleaching Conditions
Author(s) -
Kim J,
Park S,
Lee H,
Kim H,
Choi C,
Park J
Publication year - 2016
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.4956503
Subject(s) - dosimeter , irradiation , materials science , dosimetry , optically stimulated luminescence , sensitivity (control systems) , analytical chemistry (journal) , optics , nuclear medicine , radiochemistry , luminescence , chemistry , optoelectronics , physics , nuclear physics , medicine , chromatography , electronic engineering , engineering
Purpose: This work evaluated the characteristics of optically stimulated luminescence dosimeters (OSLDs) with fully filled deep electron/hole traps (OSLDfull) with the bleaching conditions according to the accumulated dose. Methods: The OSLDs were first pre‐irradiated with a Co‐60 gamma ray at more than 5 kGy, so as to fill the deep electron and hole traps. Using a 6‐MV beam, the OSLDfull characteristics were investigated in terms of the full bleaching, fading, dose linearity, and dose sensitivity obtained in response to the accumulated dose values. To facilitate a comparison of the dose sensitivity, OSLDs with un‐filled deep electron/hole traps (OSLDempty) were investigated in the same manner. A long‐pass filter was used to exclude bleaching‐source wavelengths of less than 520 nm. Various bleaching time and wavelength combinations were used in order to determine the optimal bleaching conditions for the OSLD full. Results: The fading for the OSLDfull exhibited stable signals after 8 min, for both 1‐ and 10‐Gy. For 4‐h bleaching time and an unfiltered bleaching device, the supralinear index values for the OSLDfull were 1.003, 1.002, 0.999, and 1.001 for doses of 2, 4, 7, and 10 Gy, respectively. For a 65‐Gy accumulated dose with a 5‐Gy fraction, no variation in dose sensitivity was obtained for the OSLDfull, within a standard deviation of 0.85%, whereas the OSLDempty dose sensitivity decreased by approximately 2.3% per 10 Gy. The filtered bleaching device yielded a highly stable sensitivity for OSLDfull, independent of bleaching time and within a standard deviation of 0.71%, whereas the OSLDempty dose sensitivity decreased by approximately 4.2% per 10 Gy for an accumulated dose of 25 Gy with a 5‐Gy fraction. Conclusion: Under the bleaching conditions determined in this study, clinical dosimetry with OSLDfull is highly stable, having an accuracy of 1% with no change in dose sensitivity or linearity at clinical doses. This work was supported by a National Research Foundation of Korea (NRF) grant, funded by the Korea government (MISP) (No. 2014M2B2A4031164), and by the Korea Health Technology R&D Project through the Korea Health Industry Development Institute (KHIDI), funded by the Ministry of Health & Welfare, Republic of Korea (HI14C3459).

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here