Early dislodgement of Indwelling Pleural Catheter ( IPC ): a balancing act
Author(s) -
Tung Alvin Hon Man,
Ngai Jenny Chun Li,
Ng Susanna So Shan,
Ko Fanny Wai San,
Hui David ShuCheong
Publication year - 2014
Publication title -
respirology case reports
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.304
H-Index - 9
ISSN - 2051-3380
DOI - 10.1002/rcr2.33
Subject(s) - medicine , pleural effusion , surgery , drainage , catheter , pleural fluid , effusion , ecology , biology
A 63‐year‐old nonsmoker with right malignant pleural effusion derived symptomatic benefit following drainage of his effusion. Following insertion of indwelling pleural catheter ( IPC ), 1.3 L of blood‐stained fluid was drained into underwater sealed bottle ( A trium®), but the IPC dislodged 26 h after continuous connection. We believe that the weight of the drainage bottle (including the un‐emptied fluid) and the prolonged connection time contributed to this uncommon event reported in the literature. There was no recurrence when his second IPC was connected to a drainage bag which was emptied at every 500 mL, capped at 2 h each time. An anchoring stitch should also be considered when drainage devices heavier than the manufacturer bottles are used to drain IPC .