Factors affecting signal quality in implantable cardiac monitors with long sensing vector
Author(s) -
Forleo Giovanni B.,
Amellone Claudia,
Sacchi Riccardo,
Lombardi Leonida,
Lucciola Maria Teresa,
Scotti Valentina,
Viecca Maurizio,
Schiavone Marco,
Giacopelli Daniele,
Giammaria Massimo
Publication year - 2021
Publication title -
journal of arrhythmia
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.463
H-Index - 21
eISSN - 1883-2148
pISSN - 1880-4276
DOI - 10.1002/joa3.12585
Subject(s) - medicine , confidence interval , implant , demographics , body surface area , amplitude , cardiology , surgery , optics , demography , physics , sociology
Purpose Electrical artefacts are frequent in implantable cardiac monitors (ICMs). We analyzed the subcutaneous electrogram (sECG) provided by an ICM with a long sensing vector and factors potentially affecting its quality. Methods Consecutive ICM recipients underwent a follow‐up where demographics, body mass index (BMI), implant location, and surface ECG were collected. The sECG was then analyzed in terms of R‐wave amplitude and P‐wave visibility. Results A total of 84 patients (43% female, median age 68 [58‐76] years) were enrolled at 3 sites. ICMs were positioned with intermediate inclination (n = 44, 52%), parallel (n = 35, 43%), or perpendicular (n = 5, 6%) to the sternum. The median R‐wave amplitude was 1.10 (0.72‐1.48) mV with P waves readily visible in 69.2% (95% confidence interval, CI: 57.8%‐79.2%), partially visible in 23.1% [95% CI: 14.3%‐34.0%], and never visible in 7.7% [95% CI: 2.9%‐16.0%] of patients. Men had higher R‐wave amplitudes compared to women (1.40 [0.96‐1.80] mV vs 1.00 [0.60‐1.20] mV, P  = .001), while obese people tended to have lower values (0.80 [0.62‐1.28] mV vs 1.10 [0.90‐1.50] mV, P  = .074). The P‐wave visibility reached 86.2% [95% CI: 68.3%‐96.1%] in patients with high‐voltage P waves (≥0.2 mV) at surface ECG. The sECG quality was not affected by implant site. Conclusion In ordinary clinical practice, ICMs with long sensing vector provided median R‐wave amplitude above 1 mV and reliable P‐wave visibility of nearly 70%, regardless of the position of the device. Women and obese patients showed lower but still very good signal quality.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here