0° vs. 180° CT localiser: The effect of vertical off‐centring, phantom positioning and tube voltage on dose optimisation in multidetector computed tomography
Author(s) -
AlHayek Yazan,
Zheng Xiaoming,
Davidson Rob,
Hayre Christopher,
AlMousa Dana,
Finlay Campbell,
Spuur Kelly
Publication year - 2022
Publication title -
journal of medical radiation sciences
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.484
H-Index - 18
eISSN - 2051-3909
pISSN - 2051-3895
DOI - 10.1002/jmrs.535
Subject(s) - imaging phantom , centring , supine position , nuclear medicine , medicine , materials science , biomedical engineering , surgery , engineering , mechanical engineering
Abstract Introduction Patient positioning is an essential consideration for the optimisation of radiation dose during CT examinations. The study objectives seek to explore the effects of vertical off‐centring, localiser direction (0° and 180°), and phantom positioning (supine and prone) on radiation dose, using three different tube voltages in multidetector computed tomography (MDCT) imaging. Methods The trunk of a PBU‐60 anthropomorphic phantom was imaged using a Discovery CT750 HD – 128 slice (GE Healthcare). Images employing 0° and 180° localisers were acquired in supine and prone orientation for each combination of vertical off‐centring (±100, ±60 and ±30 mm) and different tube voltages (80, 120 and 140 kVp), using the system’s automatic tube current modulation (ATCM) function. The displayed volume CT dose index (CTDI vol ) and dose length product (DLP) were recorded. Results With incremental table off‐centring of ±100 mm, the dose at 120 kVp in the supine position ranged from 63% to 196% (0° localiser) and from 66% to 191% (180° localiser) as compared to iso‐centre. While in the prone position, the dose ranged from 62% to 195% (0° localiser); and 62% to 193% (180° localiser), with a notable dose increase at higher tube voltages. Dose variation and vertical off‐centring showed a significant relationship for both 0° and 180° localisers ( r = 0.94 and 0.96, respectively, P < 0.001). The CTDI vol variation between supine and prone phantom positions at ±100 mm off‐centring was 0.22 mGy (2.9%), and 0.19 mGy (2.3%) when the 0° and 180 ° localisers were utilised, respectively. Conclusions Phantom off‐centring and localiser direction evidenced large dose variation. It is recommended that the 0° localiser is employed during CT examinations, in order to minimise the potential additional radiation dose which may result from off‐centring and the use of lower tube voltages where clinically appropriate.