Numerical simulation of ultrasonic heat meter by multiphysics coupling finite-element simulation software
Author(s) -
Lin Li,
Hongliang Zheng
Publication year - 2020
Publication title -
thermal science
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.339
H-Index - 43
eISSN - 2334-7163
pISSN - 0354-9836
DOI - 10.2298/tsci191106122l
Subject(s) - multiphysics , acoustics , ultrasonic sensor , materials science , sound pressure , coupling (piping) , finite element method , mechanics , optics , physics , metallurgy , thermodynamics
Objective: To increase heat calculation accuracy, the numerical simulation of
the ultrasonic heat meter is explored by Multiphysics coupling. Methods:
COMSOL, a Multiphysics coupling finite-element simulation software, is used
to build the coupling model of the sound field, structure field, and
electric field. The propagation of ultrasonic waves in heat meters is
simulated, and its sound field distribution in pure water is analyzed.
According to the operating conditions of ultrasonic heat meters, the
influence of impurities with different concentrations on ultrasonic
propagation is analyzed. The end-face sound pressure levels of the incident
transducer and the receiving transducer are compared to obtain the
attenuation laws of ultrasonic waves in the liquid-solid two-phase flow.
Results: The main lobe and multiple side lobes exist during the propagation
of ultrasonic waves. The energy of the main lobe is higher than that of the
side lobes. Bubbles resonate under the action of the sound field. Also,
bubbles of different diameters correspond to different resonance
frequencies, which have larger sound pressure than that of the incident
sound field. Most of the sound waves are reflected at the liquid-solid
interface, while some of them continue to propagate through the media,
affecting the sound pressure distribution on the end-face of the receiving
transducer, thereby affecting the measurement accuracy of the ultrasonic
heat meter. Conclusion: The reliability and detection efficiency of the heat
meter is improved, which is significant and theoretically valuable.