Pressure gauge calibration applying 0-A-0 pressurization to reference gauge
Author(s) -
Hiroaki Kajikawa,
Tokihiko Kobata
Publication year - 2016
Publication title -
acta imeko
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.178
H-Index - 12
eISSN - 2221-870X
pISSN - 0237-028X
DOI - 10.21014/acta_imeko.v5i1.319
Subject(s) - calibration , gauge (firearms) , cabin pressurization , pressure measurement , physics , mechanical engineering , materials science , engineering , thermodynamics , metallurgy , quantum mechanics
A calibration method is proposed to improve pressure calibration that uses a pressure gauge as the reference device. The reference gauge is pressurized by a 0-A-0 pressurization procedure, whereas the test gauge can be pressurized through various procedures. The calibration results with this method were consistent with those calibrated against a pressure balance. The method is expected to help develop low-cost pressure calibration systems that are more precise.