Automated Detection of Inaccurate and Imprecise Transitions in Peptide Quantification by Multiple Reaction Monitoring Mass Spectrometry
Author(s) -
Susan E. Abbatiello,
D.R. Mani,
Hasmik Keshishian,
Steven A. Carr
Publication year - 2009
Publication title -
clinical chemistry
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.705
H-Index - 218
eISSN - 1530-8561
pISSN - 0009-9147
DOI - 10.1373/clinchem.2009.138420
Subject(s) - analyte , selected reaction monitoring , replicate , mass spectrometry , computer science , data mining , tandem mass spectrometry , algorithm , chemistry , chromatography , mathematics , statistics
Multiple reaction monitoring mass spectrometry (MRM-MS) of peptides with stable isotope-labeled internal standards (SISs) is increasingly being used to develop quantitative assays for proteins in complex biological matrices. These assays can be highly precise and quantitative, but the frequent occurrence of interferences requires that MRM-MS data be manually reviewed, a time-intensive process subject to human error. We developed an algorithm that identifies inaccurate transition data based on the presence of interfering signal or inconsistent recovery among replicate samples.