z-logo
open-access-imgOpen Access
Comparison of M.I.C.E. and Etest with CLSI Agar Dilution for Antimicrobial Susceptibility Testing against Oxacillin-Resistant Staphylococcus spp
Author(s) -
Eloiza Helena Campana,
Cecília G Carvalhaes,
Bruonato,
Antônia Maria de Oliveira Machado,
Ana Cristina Gales
Publication year - 2014
Publication title -
plos one
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.99
H-Index - 332
ISSN - 1932-6203
DOI - 10.1371/journal.pone.0094627
Subject(s) - etest , teicoplanin , agar dilution , linezolid , vancomycin , microbiology and biotechnology , staphylococcus aureus , antimicrobial , medicine , agar , minimum inhibitory concentration , biology , bacteria , genetics
Objective The main objective of this study was to comparatively evaluate the performance of M.I.C.E. and Etest methodologies to that of agar dilution for determining the antimicrobial susceptibility profile of oxacillin-resistant Staphylococcus spp. Methods A total of 100 oxacillin-resistant Staphylococcus spp. isolates were collected from hospitalized patients at a teaching hospital. Antimicrobial susceptibility testing for vancomycin, teicoplanin and linezolid was performed using the reference CLSI agar dilution method (2009), Etest and M.I.C.E. methodologies. The MIC values were interpreted according to CLSI susceptibility breakpoints and compared by regression analysis. Results In general, the essential agreement (±1-log 2 ) between M.I.C.E. and CLSI agar dilution was 93.0%, 84.0% and 77.0% for linezolid, teicoplanin and vancomycin, respectively. Essential agreement rates between M.I.C.E. and Etest were excellent (>90.0%) for all antibiotics tested. Both strips (M.I.C.E. and Etest) yielded two very major errors for linezolid. Unacceptable minor rates were observed for teicoplanin against CoNS and for vancomycin against S. aureus . Conclusions According to our results, linezolid and teicoplanin MICs against all staphylococci and S. aureus , respectively, were more accurately predicted by M.I.C.E. strips. However, the Etest showed better performance than M.I.C.E. for predicting vancomycin MICs against all staphylococci. Thus, microbiologists must be aware of the different performance of commercially available gradient strips against staphylococci.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom