Comparison of the Finapres Blood Pressure Monitor with Intra-arterial Manometry during Induction of Anaesthesia
Author(s) -
J. L. Kermode,
Naomi Davis,
William R. Thompson
Publication year - 1989
Publication title -
anaesthesia and intensive care
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.494
H-Index - 62
eISSN - 1448-0271
pISSN - 0310-057X
DOI - 10.1177/0310057x8901700413
Subject(s) - medicine , anesthesia , blood pressure , plethysmograph , cuff , continuous noninvasive arterial pressure , surgery
The Finapres (Ohmeda, Madison, U.S.A.) is a non-invasive device which continuously measures the arterial blood pressure in a finger and produces a real-time display of the arterial pressure wave. It consists of a finger cuff with an infra-red transmission plethysmograph, a servo control box and a monitor unit. The device was compared with intra-arterial pressure monitoring in twenty patients during induction of anaesthesia for elective neurosurgical procedures. The differences between the two methods were considerable, ranging from −40 mmHg to +26 mmHg for mean pressure. While the Finapres has potential as a non-invasive continuous blood pressure monitor, the current model Finapres, as supplied, displays too great a variability for it to be used as an alternative to intra-arterial pressure monitoring.