Determination of Aluminium in Serum: Findings of an External Quality Assessment Scheme
Author(s) -
Andrew Taylor,
B J Starkey,
A. W. Walker
Publication year - 1985
Publication title -
annals of clinical biochemistry international journal of laboratory medicine
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.6
H-Index - 80
eISSN - 1758-1001
pISSN - 0004-5632
DOI - 10.1177/000456328502200403
Subject(s) - external quality assessment , atomic absorption spectroscopy , aluminium , scheme (mathematics) , chemistry , mathematics , medicine , materials science , metallurgy , physics , pathology , quantum mechanics , mathematical analysis
The analytical performance of laboratories participating in the serum aluminium programme of the Guildford Trace Element Quality Assessment Scheme has been evaluated. The considerable disparity between reported results was not attributable to the nature of the samples distributed, nor to the standard aluminium solutions used in different laboratories. All laboratories used electrothermal atomic absorption spectrometry but sample preparation procedures and temperature programmes varied enormously. While some laboratories performed better than others in the scheme, and the standard of analysis appears to be improving with time, no laboratory showed consistent acceptable performance.