z-logo
open-access-imgOpen Access
A Continuous Cardiac Output Recorder Employing the Fick Principle
Author(s) -
Arthur C. Guyton,
FARISH CHARLES A.,
ABERNATHY J. BERRY
Publication year - 1959
Publication title -
circulation research
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 4.899
H-Index - 336
eISSN - 1524-4571
pISSN - 0009-7330
DOI - 10.1161/01.res.7.4.661
Subject(s) - cardiac output , fick principle , arteriovenous oxygen difference , calibration , oxygen , heart rate , biomedical engineering , mathematics , cardiology , medicine , blood pressure , statistics , chemistry , hemodynamics , stroke volume , organic chemistry
A continuous cardiac output recorder has been developed utilizing an oxygen consumption recorder, an arteriovenous oxygen difference recorder and a computer. The computer continuously derives and records the cardiac output by dividing the rate of oxygen consumption by the A-V oxygen difference. Calibration of the continuous cardiac output recorder against a Ludwig stromuhr showed that under many different conditions and many different levels of cardiac output the 305 individual calibration points had a mean deviation from the regression line of ± 7.7 per cent. This correlation is as satisfactory as previously reported correlations between the intravenous dye method for recording cardiac output and the direct Fick method.