Myocardial Necrosis from Direct Current Countershock
Author(s) -
Charles F. Dahl,
Gordon A. Ewy,
E. D. Warner,
E.D. Thomas
Publication year - 1974
Publication title -
circulation
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 7.795
H-Index - 607
eISSN - 1524-4539
pISSN - 0009-7322
DOI - 10.1161/01.cir.50.5.956
Subject(s) - paddle , medicine , electrode , necrosis , anesthesia , cardiology , surgery , materials science , composite material , chemistry
The effect of varying both paddle electrode size and the time interval between direct current countershock on myocardial necrosis was studied. Forty-two dogs were divided into seven groups of six dogs each. All dogs were given ten consecutive, 240 watt-second countershocks (delivered energy into a 50 ohm load). Three groups were shocked with paddle electrode diameters of 8.0 cm (standard electrodes), two groups with paddle electrode diameters of 12.8 cm (large electrodes), and two groups with paddle electrode diameters of 4.3 cm (small electrodes). The time intervals between discharges in the groups shocked with the standard electrodes were 15 seconds, one minute, and three minutes. The time interval between discharges in the groups shocked with small and large electrodes was 15 seconds and three minutes. Myocardial necrosis was quantitated by precordial electrocardiographic mapping recorded minutes after, and by gross and microscopic examination of the hearts four days after direct current countershock.When the time interval between discharges was shorter, myocardial necrosis was greater. When the time interval between discharges was constant, more necrosis was produced with smaller-sized paddle electrodes. It is concluded that large paddle electrodes should be used for delivering direct current countershocks, and that during elective cardioversion, consecutive discharges should be delivered at time intervals greater than three minutes.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom