Twiddler's syndrome in a patient with a biventricular-defibrillator device
Author(s) -
Faizel Osman,
Ross W. Ward,
A. Morley-Davies
Publication year - 2008
Publication title -
ep europace
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 2.119
H-Index - 102
eISSN - 1532-2092
pISSN - 1099-5129
DOI - 10.1093/europace/eum292
Subject(s) - medicine , cardiology , cardiac resynchronization therapy , myocardial infarction , heart failure , surgery , ejection fraction
A 69-year-old man with previous myocardial infarction and proximal three-vessel coronary artery disease underwent coronary bypass grafting, with an epicardial lead placed on the lateral left ventricular wall during surgery. A cardiac resynchronization therapy-defibrillator (CRT-D) device was subsequently implanted using active right atrial and right ventricular leads, with the pulse generator placed in a pre-pectoral pocket. Four weeks later, the right atrial lead was failing to sense or capture, and chest X-ray revealed it had pulled out of the myocardium and coiled up behind the device; a diagnosis of Twiddler's syndrome was made. Twiddler's syndrome is unusual in patients with CRT-D devices and may cause symptoms such as inappropriate shocks and hiccups. Placement of the pulse generator in a sub-pectoral position may help prevent it.