Determination of Serum Calcium by Use of an AutoAnalyzer I-II Hybrid
Author(s) -
C. Jacklyn,
C. Daniel Freeman,
Mike Gray,
Malcolm Macaulay
Publication year - 1973
Publication title -
clinical chemistry
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.705
H-Index - 218
eISSN - 1530-8561
pISSN - 0009-9147
DOI - 10.1093/clinchem/19.5.516
Subject(s) - autoanalyzer , sma* , calcium , chemistry , chromatography , continuous flow , computer science , engineering , biochemical engineering , algorithm , organic chemistry
A hybridized system is described for the continuous-flow determination of serum calcium. The analysis is based on the cresolphthalein colorimetric procedure as described for use with the SMA 12/60 (Technicon Corp.). The hydraulic features of the AutoAnalyzer II are used with modules of the AutoAnalyzer I. This combination gives both the ease of operation and hydraulic improvements inherent in the AutoAnalyzer II system and provides an emergency alternative to the SMA 12/60, with which results are shown to be comparable.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom