Determination of Cr, Fe, Co, Ni, Cu, Zn, As and Pb in liquid chemical waste by energy dispersive X-ray fluorescence
Author(s) -
Eduardo de Almeida,
Glauco Arnold Tavares,
José Albertino Bendassolli,
V. F. Nascimento Filho
Publication year - 2010
Publication title -
journal of radioanalytical and nuclear chemistry
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.374
H-Index - 67
eISSN - 1588-2780
pISSN - 0236-5731
DOI - 10.1007/s10967-010-0694-2
Subject(s) - detection limit , x ray fluorescence , analytical chemistry (journal) , fluorescence , materials science , spectrometer , filter paper , chemistry , chromatography , optics , physics
A method for simultaneous determination of Cr, Fe, Co, Ni, Cu, Zn, As e Pb in liquid chemical waste using Energy Dispersive X-Ray Fluorescence (EDXRF) technique was evaluated. A small sample amount (200 mu L) was dried on a 6.35 mu m thickness Mylar film at 60 degrees C and the analyses were carried out using an EDXRF spectrometer operated with an X-ray Mo tube (Zr filter) at 30 kV/20 mA. The acquisition time was 300 s and the Ga element was utilized as internal standard at 25 mg/L for quantitative analysis. The method trueness was assessed by spiking and the detection limit for those elements ranged from 0.39 to 1.7 mg/L. This method is notable because it assists the choice of the more appropriated waste treatment procedure, in which inter elemental interference is a matter of importance. In addition, this inexpensive method allows a non-destructive determination of the elements from (19)K to (92)U simultaneously
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom