Subject(s) - langmuir probe , plasma , electron density , arc (geometry) , atmospheric pressure , atomic physics , vacuum chamber , electron temperature , range (aeronautics) , plasma diagnostics , analytical chemistry (journal) , chemistry , atmospheric pressure plasma , electric arc , electron , plasma parameters , materials science , electrode , physics , chromatography , nuclear physics , meteorology , mathematics , geometry , composite material
In this piece of study, the design and development of a low cost vacuum chamber with different ports are mentioned. I-V characteristic of Langmuir Probe is obtained using the primary data of the experimental set up for ‘single probe method’ in nitrogen seeded arc plasma at atmospheric pressure The floating potential was found to be 36V and the electron density to be 1.24 ´ 104 K. Variation in electron density of the arc plasma with probe potential is also studied at low pressure range from 10 Nm-2 (0.10 mbar) to 20 Nm-2.BIBECHANA 16 (2019) 1-6