Unusual Wenckebach Upper Rate Response of an Atrial‐Based DDD Pacemaker
Author(s) -
FREDMAN CAREY S.,
BJERREGAARD PREBEN,
JANOSIK DENISE L.,
BAROLD S. SERGE
Publication year - 1992
Publication title -
pacing and clinical electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.686
H-Index - 101
eISSN - 1540-8159
pISSN - 0147-8389
DOI - 10.1111/j.1540-8159.1992.tb03089.x
Subject(s) - medicine , cardiology , prolongation , ventricular rate , p wave , heart rate , anesthesia , atrial fibrillation , blood pressure
We describe in this report an unusual form of Wenckebach upper rate response produced by a DDD pulse generator with atrial‐based lower rate timing. The pacemaker maintained the programmed upper and lower rate intervals at the expense of a prolonged atrial paced‐venfricular paced AV interval. This form of upper rate behavior eliminated the longer cycle (containing the unsensed P wave) that occurs at the end of the pacemaker Wenckebach sequence during traditional DDD pacing with ventricular‐based lower rate timing.