z-logo
Premium
Analytical and clinical performance of a new, automated assay panel for the diagnosis of antiphospholipid syndrome
Author(s) -
DE MOERLOOSE P.,
REBER G.,
MUSIAL J.,
ARNOUT J.
Publication year - 2010
Publication title -
journal of thrombosis and haemostasis
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.947
H-Index - 178
eISSN - 1538-7836
pISSN - 1538-7933
DOI - 10.1111/j.1538-7836.2010.03857.x
Subject(s) - antiphospholipid syndrome , medicine , intensive care medicine , thrombosis
Summary.  Background:  Anticardiolipin (aCL) and anti‐β 2 glycoprotein I (aβ 2 GPI) antibodies are part of the criteria for antiphospholipid syndrome (APS). Therefore they are widely measured and about 30 commercial kits are available. Objectives:  To investigate the analytical and clinical performances of four fully automated, chemiluminescent assays: HemosIL AcuStar aCL IgG, HemosIL AcuStar aCL IgM, HemosIL AcuStar aβ 2 GPI IgG, and HemosIL AcuStar aβ 2 GPI IgM. Methods:  Cut‐off values were assessed by testing 250 blood donors. Total coefficients of variation (CV) were determined with six plasma pools and two controls ranging from 4.3 to 2694 U mL −1 depending on the assay. Samples from 218 well‐characterized patients and 103 controls were measured in three laboratories to determine inter‐laboratory variation. The results of the 321 samples were compared with three commercial assays (REAADS, INOVA and VARELISA). Results:  Cut‐off values were assigned to 20 arbitrary units for all the tests. Total CV ranged from 4.3 to 11.2%. No interference of hemoglobin, bilirubin, triglycerides, heparins and rheumatoid factor was observed. Inter‐laboratory variability was low and no sample changed status. Overall status agreement between HemosIL assays and the comparator kits ranged from 82 to 96%. Sensitivity, specificity, agreement when predicting APS and the odds ratios when predicting a thrombotic or obstetric event gave comparable results between HemosIL AcuStar and the three other assays. Conclusions:  Our study demonstrates that the fully automated HemosIL AcuStar aPL assay panel showed similar performances to the three commercial ELISAs commonly used by various laboratories to detect antiphospholipid antibodies.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here