z-logo
Premium
A laboratory study of four electric pulp testers
Author(s) -
DUMMER P.M.H.,
TANNER M.,
McCARTHY J.P.
Publication year - 1986
Publication title -
international endodontic journal
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.988
H-Index - 119
eISSN - 1365-2591
pISSN - 0143-2885
DOI - 10.1111/j.1365-2591.1986.tb00472.x
Subject(s) - voltage , pulp (tooth) , constant voltage , materials science , electrical engineering , environmental science , engineering , dentistry , medicine
Summary. Four commercially available, battery operated, unipolar electric pulp testers were tested and compared under identical laboratory conditions. The form and duration of the electrical impulses varied considerably between the instruments as did the maximum output voltage. None of the models tested passed constant current values when connected to a range of resistances.