z-logo
Premium
Calibration of a Floscope peak flow meter and a Vitalograph pulmonary monitor
Author(s) -
FISHER J.,
SHAW A.
Publication year - 1981
Publication title -
anaesthesia
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.839
H-Index - 117
eISSN - 1365-2044
pISSN - 0003-2409
DOI - 10.1111/j.1365-2044.1981.tb08863.x
Subject(s) - medicine , repeatability , calibration , metre , flow measurement , peak flow meter , statistics , mathematics , mechanics , physics , astronomy , asthma
Summary The performance of a Floscope peak flow meter and the Vitalograph pulmonary monitor are assessed using an absolute calibration method. Both instruments have good repeatability. For peak flows greater than 150 litres/minute, the Floscope showed errors of between −12.4% and +7.1%. For flows below 150 litres/minute the percentage errors were greater. The pulmonary monitor showed errors of between −20% and −5.5%. Its high resistance makes it an inappropriate instrument to use for absolute measurement of peak flows. It is useful for assessing trends.