A powder diffractometer for a synchrotron source
Author(s) -
Sabine T. M.
Publication year - 1987
Publication title -
journal of applied crystallography
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.429
H-Index - 162
ISSN - 1600-5767
DOI - 10.1107/s0021889887086886
Subject(s) - monochromator , diffractometer , synchrotron , analyser , optics , powder diffractometer , materials science , synchrotron radiation , physics , diffraction , wavelength , scanning electron microscope
A design is given for a powder diffractometer on a synchrotron source. A reasonable balance between resolution and luminosity is achieved by making the effective mosaic spread parameters of the monochromator and analyser crystal equal to the vertical divergence of the synchrotron beam. A monochromator with a Bragg angle of 60° should be used. Care is necessary in preparing specimens for data collection to avoid the hazards of particle‐size broadening on one hand and extinction on the other. The conclusions are compared with recent experimental work.