Quantitative Determination of n-Propane, iso-Butane, and n-Butane by Headspace GC-MS in Intoxications by Inhalation of Lighter Fluid*
Author(s) -
MariePaule Bouche,
Willy E. Lambert,
Jan F. P. Van Bocxlaer,
M Piette,
A.P. De Leenheer
Publication year - 2002
Publication title -
journal of analytical toxicology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.161
H-Index - 76
eISSN - 1945-2403
pISSN - 0146-4760
DOI - 10.1093/jat/26.1.35
Subject(s) - butane , chromatography , propane , gas chromatography , chemistry , tenax , extraction (chemistry) , sample preparation , mass spectrometry , analytical chemistry (journal) , organic chemistry , catalysis
This report describes a fully elaborated and validated method for quantitation of the hydrocarbons n-propane, iso-butane, and n-butane in blood samples. The newly developed analytical procedure is suitable for both emergency cases and forensic medicine investigations. Its practical applicability is illustrated with a forensic blood sample after acute inhalative intoxication with lighter fluid; case history and toxicological findings are included. Identification and quantitation of the analytes were performed using static headspace extraction combined with gas chromatography-mass spectrometry. In order to reconcile the large gas volumes injected (0.5 mL) with the narrowbore capillary column and thus achieve preconcentration, cold trapping on a Tenax sorbent followed by flash desorption was applied. Adequate retention and separation were achieved isothermally at 35 degrees C on a thick-film capillary column. Sample preparation was kept to a strict minimum and involved simply adding 2.5 microL of a liquid solution of 1,1,2-trichlorotrifluoroethane in t-butyl-methylether as an internal standard to aliquots of blood in a capped vial. Standards were created by volumetric dilution departing from a gravimetrically prepared calibration gas mixture composed of 0.3% of n-propane, 0.7% of iso-butane, and 0.8% of n-butane in nitrogen. In the forensic blood sample, the following concentrations were measured: 90.0 microg/L for n-propane, 246 microg/L for iso-butane, and 846 microg/L for n-butane.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom