Quantitative gas chromatography, using retention times
Author(s) -
Brandt A. E.,
Lands W. E. M.
Publication year - 1968
Publication title -
lipids
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.601
H-Index - 120
eISSN - 1558-9307
pISSN - 0024-4201
DOI - 10.1007/bf02531737
Subject(s) - retention time , chromatography , sample (material) , gas chromatography , column (typography) , analytical chemistry (journal) , diffusion , chemistry , point (geometry) , mathematics , physics , geometry , connection (principal bundle) , thermodynamics
Abstract Diffusion of an injected sample within a gas chromatographic column does not begin from a point source but from a band. Therefore the method of calculating relative areas by using retention time × peak height may require a correction factor to give a more accurate estimate of peak areas. When this correction was applied, the analysis was comparable with that obtained by the more time‐consuming triangulation method.