Simultaneous in vivo comparison of water‐filled and air‐filled pressure measurement catheters: Implications for good urodynamic practice
Author(s) -
Gammie A.,
Abrams P.,
Bevan W.,
EllisJones J.,
Gray J.,
Hassine A.,
Williams J.,
Hashim H.
Publication year - 2016
Publication title -
neurourology and urodynamics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.918
H-Index - 90
eISSN - 1520-6777
pISSN - 0733-2467
DOI - 10.1002/nau.22827
Subject(s) - medicine , pressure measurement , mechanical engineering , engineering
Aims This study aimed to evaluate whether the pressure readings obtained from air‐filled catheters (AFCs) are the same as the readings from simultaneously inserted water‐filled catheters (WFCs). It also aimed to make any possible recommendations for the use of AFCs to conform to International Continence Society (ICS) Good Urodynamic Practices (GUP). Methods Female patients undergoing urodynamic studies in a single center had water‐filled and air‐filled catheters simultaneously measuring abdominal and intravesical pressure during filling with saline and during voiding. The pressures recorded by each system at each event during the test were compared using paired t‐test and Bland‐Altman analyses. Results 62 patients were recruited, of whom 51 had pressures that could be compared during filling, and 23 during voiding. On average, the pressures measured by the two systems were not significantly different during filling and at maximum flow, but the values for a given patient were found to differ by up to 10 cmH 2 O. Conclusions This study shows that AFCs and WFCs cannot be assumed to register equal values of pressure. It has further shown that even when the p det readings are compared with their value at the start of a test, a divergence of values of up to 10 cmH 2 O remains. If AFCs are used, care must be taken to compensate for any p det variations that occur during patient movement. Before AFCs are adopted, new normal values for resting pressures need to be developed to allow good quality AFC pressure readings to be made. Neurourol. Urodynam. 35:926–933, 2016 . © 2015 Wiley Periodicals, Inc.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom